You are here
Multiwell-MEA-System
Multiwell-MEA workstation for recording and analyzing data from multiwell MEA plates. With integrated data acquisition for recording from 288 MEA electrodes and 8 additional analog channels. Integrated filter amplifier and voltage stimulus generator (simultaneous stimulation in all wells possible. Any electrode can be selected for stimulation).
Complete with Multiwell-MEA headstage, interface board, and accessories. USB High Speed data transfer with a sampling rate of up to 50 kHz/channel.
• Multiwell-MEA headstage:
 Pre- and filter amplifier with integrated data acquisition and
   analog-digital converter
 - Bandwidth adjustable via software
 - Data resolution 24 bit
 - Direct access to each electrode for stimulation via internal
   stimulus generator.
 - Control of stimulator via the included data acquisition
   software Multiwell Screen
 - Integrated blanking circuit for stimulus artifact suppression
 - Integrated stimulus generators (any electrode can be selected
   for stimulation)
 - Current stimulation at max. +/-1 mA
 - Voltage stimulation at max. +/-10 V
 - Sampling rate max. 50 kHz per channel
 - Integrated heating plate and temperature control
• Interface board multiboot (IFB-C):
 Interface board with signal processor for real-time signal
 detection and feedback
 Includes various in-/outputs:
 - 16 possible boot images to operate all MCS headstages
of the modular 2100 amplifier suite
 - 16-bit digital input/output channels for TTL integration, 3.3/5 V compatible
(All channels accessible on the backside via Honda-PCS-XE68LFD,
additional bit 0-3 digital in and bit 0-3 digital out accessible on
the front via LEMO EPL.00.250.NTN)
 - Channel definition via software: IFB-Control
(free to download) – Boot image definition
 - 8 analog input channels, +/- 10 V voltage input range
@ 24 bit ADC accessible on the front
(All channels via DIL10Header-100mil, channel 1 and 2
additional via LEMO EPL.00.250.NTN)
 - Audio/stereo out (standard stereo jack 3.5 mm)
 - USB-3.0 type C data transfer to PC (cable included)
 - iX industrial, type B cable for connecting IFB-C to
headstage/signal collector unit/receiver (cable included)
 - DSP-JTAG for direct programming of the built-in DSP
(DIL14Header-100mil-angeled)
 - Physical ground socket (4 mm socket)
 - External power supply included: 24 V/65 W, 110-240 V
• Multiwell Screen program:
 - Data acquisition and analysis software
 - Including a 2 years maintenance package (software updates
   and support)
• Accessories:
 - USB memory stick with user manuals and software
Technical specifications as specified in the data sheet unless noted otherwise in the order.